Laser-polarized Xe-129 NMR and MRI at ultralow magnetic fields

TitleLaser-polarized Xe-129 NMR and MRI at ultralow magnetic fields
Publication TypeJournal Article
Year of Publication2002
AuthorsWong-Foy A, Saxena S, Moule A.J, Bitter H.ML, Seeley J.A, McDermott R, Clarke J, Pines A
JournalJournal of Magnetic Resonance
Volume157
Issue2
Pagination235-241
Date PublishedAug
ISBN Number1090-7807
Accession NumberWOS:000178377400009
Keywordsatoms
Abstract

Laser-polarized Xe-129 and a high-T-c superconducting quantum interference device (SQUID) are used to obtain magnetic resonance images in porous materials at a magnetic field of 2.3 mT, corresponding to a Larmor frequency of 27 kHz. Image resolution of 1 mm is obtained with gradients of only 1 mT/m. The resolution of xenon chemical shifts in different physicochemical environments at ultralow fields is also demonstrated. Details of the circulating flow optical pumping apparatus and the SQUID spectrometer are presented. (C) 2002 Elsevier Science (USA).

URL<Go to ISI>://WOS:000178377400009
DOI10.1006/Jmre.2002.2592
Short TitleLaser-polarized Xe-129 NMR and MRI at ultralow magnetic fields
Download Publication: